0

Transmitters for Moisture Content in Oil

Transmitters for Moisture Content in Oil
Transmitters for Moisture Content in Oil
Price And Quantity
  • 1
  • Piece/Pieces
  • 50000.00 - 150000.00 INR
Product Specifications
  • + 0.01-+ 0.02 %
  • Black, Silver and Grey
  • 15-145 PSI
  • Metal and Plastic
  • For Moisture Content In Oil
  • 8-35 Volt (v)
    Product Description

    Owing to rich field knowledge as well as professional workforce, we have achieved a noted position as a leading exporter and importer of Transmitters for Moisture Content in Oil. Provided transmitter finds its application in several industries for detection of moisture in insulation and lubrication oil. Accessible in variety of technical specifications to meet patron’s specific requirements and needs, these transmitters for moisture content in oil are developed using sophisticated technology in tune with set international parameters. 

    Technical Specification :

    Water activity sensor1)

    HC1000-400

    Measuring range1)

    0...1aw

    Accuracy incl. hysteresis and nonlinearity in air
    - Special calibration against certified standards
    - Standard calibration

    + 0.01aw (0...0.9aw) + 0.02aw (0.9...1aw)
    +
    0.02aw (0...0.9aw) + 0.03aw (0.9...1aw)

    Temperature dependence of electronics

    typ. + 0.0001 [1/oC] (typ. + 5.6 * 10-5 [1/oF])

    Temperature dependence of sensing probe

    typ. + (0.00002 + 0.0002 x aw) x ∆T [oC] ∆T = T - 20oC

    Response time with stainless steel filter at 20oC / t90

    typ. 10min in still oil

    Temperature

    Temperature sensor element

    Pt1000 (tolerance class A, DIN EN 60751)

    Working range sensing probe

    -40...180oC (-40...356oF)

    Temperature dependence of electronics

    typ. + 0.005oC/oC

    Outputs


    Two freely selectable and scaleable analogue outputs

    0 - 5V
    0 - 10V
    4 - 20mA
    0 - 20mA

    -1mA < IL < 1mA
    -1mA < IL < 1mA
    RL < 500 Ohm
    RL < 500 Ohm

    Adjustable measurement range


    Water activity

    from up to units
             aw   01

    Temperature

    T -40 (-40) 180 (356) oC (oF)

    Water content 3)

    x 0 1000 ppm

    General

    Supply voltage

    SELV 8...35V DC SELV = Safety Extra Low Voltage
    SELV 12...30V AC (optional 100...240V AC, 50/60Hz)

    Current consumption :
    - 2x voltage output
    - 2x current output

    for 24V DC/AC:

    typ. 40mA
    typ. 80mA

    Pressure range sensing pobe

    0.01...10bar (0.15...145psi)

    System requirements for software

    WINDOWS 98 or later; serial interface

    Serial interface for configuration 4)

    RS232C

    Housing / Protection class

    PC or Al Si 9 Cu 3 / IP65; Nema 4

    Cable gland

    M16 x 1.5 cable Ø 4.5 - 10 mm (0.18 - 0.39")

    Electrical connection

    screw terminals up to max. 1.5mm² (AWG 16)

    Sensor protection

    stainless steel filter

    Operating temperature range of electronics

    -40...60oC (-40...140oF)

    Working and storage temperature range Housing with display

    -20...50oC (-4...122oF)

    Storage temperature

    -40...60oC (-40...140oF)

    Electromagnetic compatibility according to

    EN61000-6-2 EN 61000-6-3

    ICES-003 ClassB

    EN 61326-1+A1+A2

    FCC Part15 ClassB

    Options

    Display

    graphical LCD (128x32 pixels), with integrated push- buttons for selecting parameters and MIN/MAX function

    Alarm outputs

    2 x 1 switch contact: 250V AC / 6A and 28V DC / 6A threshold + hysteresis can be adjusted with configuration software

    Switching parameters (freely selectable)

    aw Water activity
    T Temperature
    x Water content

    Trade Information
    • Cash Advance (CA), Cash in Advance (CID)
    • 100 Per Week
    • 3 Days
    • Yes
    • Sample costs shipping and taxes has to be paid by the buyer
    • All India
      Contact Us

      401 & G-414-416 A Wing, Kailash Indl. Complex, Near Dynalog, Behind Godrej Colony, Park Site, Vikhroli (West), Mumbai, Maharashtra, 400079, India
      Phone :+918068210843